| Make | SHK Instrument Solution |
| Model | UCDBTC-90 |
| Range | -90 to 25 °C |
| Resolution | 0.1 °C |
| Control Accuracy | ± 0.1 °C |
| Thermal In Stability | ±0.020 °C |
| Control Accuracy | ± 0.1 °C | Display | LCD |
| Hysteresis | ± 0.05 °C | Control Method | Self tuned PID controller |
| Stabilization Time | 10 minutes after set point is achieved | Size (H × W × D) | 520(H) x 250(W) x 250(D) mm |
| Thermal Instability | ±0.020 °C at -90°C, ±0.016 °C at -50°C, ±0.011 °C at 25°C | Weight | 108 Kg |
| Thermal Non-Uniformity | ±0.070 °C at -90°C, ±0.052 °C at -50°C, ±0.021 °C at 25°C | Well Immersion Depth | 120 mm |
| Cooling Time | 120 Min (Ambient to -90°C) @25°C | Insert OD | 28 mm |
| Heating Time | 15 Min (-100°C to Ambient) @25°C | Insert Length | 120 mm |
| Power Supply | 230 V(AC), 50 Hz | Operating Conditions | 5°C … 25°C |
| Power Consumption | 300 Watt | Color | Orange |
The voltage requirement is 230 V AC, 50 Hz.
The weight of the calibrator is 108 Kg.
The temperature range is -90°C to 25°C.
Cooling time is 120 minutes (Ambient to -90°C) @25°C.
SHK Instrument Solution is a trusted manufacturer and supplier of advanced calibration instruments based in Indore, India.