| Make | SHK Instrument Solution |
| Model | HIRTC500 |
| Range | 50 to 500 °C |
| Resolution | 0.1 °C |
| Control Accuracy | ± 0.1 °C |
| Thermal In Stability | ±0.09 °C |
| Control Accuracy | ± 0.1 °C | Display | LCD |
| Hysteresis | ± 0.09 °C | Control Method | Self-tuned PID controller |
| Stabilization Time/settling time | 10 minutes after set point is achieved | Size (H × W × D) | 120(H) x 200(W) x 250(D) mm |
| Thermal Instability | ±0.09 °C at 50°C, ±0.17 °C at 300°C, ±0.20 °C at 500°C | Weight | 5 Kg |
| Thermal Non-Uniformity Radial | ±0.10 °C at 50°C, ±0.19 °C at 300°C, ±0.25 °C at 500°C | Aperture Emissivity | 0.93 ± 0.02 |
| Cooling Time | 30 Min (500°C to Ambient) @25°C | Aperture Diameter | 50 mm |
| Heating Time | 10 Min (Ambient to 500°C) @25°C | Operating Conditions | 5°C … 25°C |
| Power Supply | 230 V(AC), 50 Hz | Power Consumption | 300 Watt |
The voltage requirement is 230 V AC, 50 Hz.
The weight of the calibrator is 5 Kg.
The aperture emissivity is 0.93 ± 0.02 and the aperture diameter is 50 mm.
Cooling time is 30 minutes (500°C to Ambient) @25°C and heating time is 10 minutes (Ambient to 500°C) @25°C.]de
SHK Instrument Solution is a trusted manufacturer and supplier of advanced calibration instruments based in Indore, India.